Embla S7000 Specifications |
|
| |
|
|
Bedside Unit |
|
| Signal
Acquisition Channels |
12
referential channels
8 bipolar channels
1 light sensor channel
1 event signal |
| Bipolar
Channel Characteristics |
Input Range:
+-350mV
Sampling Rates: Software selectable
for each channel between 2, 10, 20,
50, 100, 200, 500 and 2000Hz.
Low Cut Filter: Software selectable
by channel.
Frequency Response: Max 0-90Hz
based on sampling rate
Signal Noise: <2u VRms at
200Hz |
| Referential
Channel Characteristics |
Input Range:
+-75mV dynamic range
Sampling Rates: Software selectable
for each channel between 50, 100,
200Hz.
Low Cut Filter: Fixed DC at
0.3Hz & software selectable by
channel.
Frequency Response: Max 0.3-90Hz
based on sampling rate
Signal Noise: <1u VRms at
200Hz |
| Input
Impedence |
>20 Mohm |
| Automatic
Electrode Impedance Check |
Impedance test output
waveform is 0.25 µA square wave,
20-30Hz |
| Power-Line
Notch Filters |
Software selectable
for each channel. If activated a notch
filter will reject the 50 or 60Hz
power line frequencies when sampling
at a higher sampling rate than 100Hz |
| Dimension |
Height 27mm
(1.06")
Width 130mm (5.12")
Length 190mm (7.48") |
| Weight |
620g (1.37lbs) |
|
|
| |
|
| |
|
|
Patient Unit |
|
| Signal
Acquisition Channels |
4 External Channels:
The external channels are supported
by the proxy.
Signals supported by X10 proxy:
 |
|
Thermistor |
 |
|
Snore sensor |
| |
|
(vibration) |
 |
|
Abdomen belt & Thorax
belt |
| |
|
(XactTrace belts) |
 |
|
Nonin oximeter |
 |
|
|
SpO2 averaged & beat
to beat |
 |
|
|
Pulse |
 |
|
|
Pulse waveform |
| |
|
|
(plethysmograph) |
 |
|
|
Quality |
3 Internal Channels:
Pressure signal - nasal cannula
or CPAP pressure
Pressure Range:
 |
|
Nasal Cannula +- 9mm H20 |
 |
|
CPAP +- 30 mm H20 |
Integrated gravity based sensor that
provides:
 |
|
Activity |
 |
|
Body Position |
|
| Dimension |
Height:
18.5mm (0.73")
Width: 80mm (3.15")
Length: 110mm (4.33") |
| Weight |
280g (0.62lbs) |
|
|
| |
|
| |
|
|
Communication Unit |
|
| High-Speed
Digital Port |
General-purpose
digital port capable of communicating
with other components, such as the
Patient and Bedside units. |
| Serial
Ports |
Two
isolated serial ports for external
device communications
with RS232
port speed up to 115.2 k bits per
second |
| Auxiliary
Channel Characteristics |
8
isolated AUX input channels
Input range: ±5 V
Sampling rates: Software selectable
for each channel, between 2, 10, 20,
50, 100 and200Hz or 2, 16, 32, 64,
128 and 256Hz
Low Cut Filter: Software selectable
frequencies for each channel |
| 10base-T
LAN Connection |
Direct
connection to an Ethernet based local
area network |
| Power
Supply |
Built-in
medical grade power supply with
high fluctuation tolerance
Isolation:
Reinforced isolation (EN60601-1
type BF)
Classification: Class II
AC mains voltage: 115V/230V
Power supply ratings: 160mA
|
| Dimension |
Height:
80mm (3.15")
Width: 200mm (7.87")
Length: 290mm (11.42") |
| Weight |
2300g (4.41lbs) |
|
|
| |
|
| |
|
|
|
|
| |
|
|
|
|